A Bio-Impedance Readout IC With Complex-Domain Noise-Correlated Baseline Cancellation

Haidam Choi, Song I. Cheon, Gichan Yun, Sein Oh, Ji Hoon Suh, Sohmyung Ha, Minkyu Je

Research output: Contribution to journalArticlepeer-review

Abstract

This article presents a bio-impedance (bioZ) readout IC featuring a complex-domain noise-correlated baseline cancellation to overcome the limitation of the conventional baseline cancellation with the real-domain noise correlation. The proposed technique is especially beneficial in cases with significant phase shifts between the excitation current from the current generator (CG) and the input voltage signal produced across the bioZ, through which the CG current flows. This technique employs a tunable reference impedance (TRI) and adaptively adjusts it to match the bioZ in both magnitude and phase, thereby achieving a complex-domain correlation of CG noise. By flowing an identical CG current through both the TRI and bioZ, the noise voltages across the TRI and bioZ caused by the CG current become closely correlated with each other, even in the presence of substantial phase shifts, enabling effective CG noise removal after baseline subtraction. Furthermore, this work proposes a differential-difference current-balancing instrumentation amplifier (DD-CBIA) with quiet chopping for baseline subtraction, offering low power consumption, wide input range, and low input-dependent noise. Measurement results demonstrate significant enhancements in noise performance by a factor of 2.47 and 4.88 for the bioZs with phases of 30° and 60°, respectively, achieving a signal-to-noise ratio (SNR) of 101.5 dB and a figure of merit (FoM) of 150.0 dB. Validation through human-subject experiments using two-electrode configurations on the chest and wrist further supports the effectiveness of the proposed bioZ readout IC.

Original languageEnglish (US)
Pages (from-to)3538-3548
Number of pages11
JournalIEEE Journal of Solid-State Circuits
Volume59
Issue number11
DOIs
StatePublished - 2024

Keywords

  • Baseline cancellation
  • bio-impedance (bioZ)
  • current generator (CG)
  • impedance spectroscopy
  • instrumentation amplifier (IA)
  • low noise
  • tunable reference impedance (TRI)
  • two-electrode measurement
  • wearable device

ASJC Scopus subject areas

  • Electrical and Electronic Engineering

Fingerprint

Dive into the research topics of 'A Bio-Impedance Readout IC With Complex-Domain Noise-Correlated Baseline Cancellation'. Together they form a unique fingerprint.

Cite this