A design method of organ-on-a-chip with highly accurate measurement of trans-epithelial electrical resistance

Takashi Miyazaki, Yoshikazu Hirai, Ken Ichiro Kamei, Toshiyuki Tsuchiya, Osamu Tabata

Research output: Contribution to journalArticlepeer-review

Abstract

Trans-epithelial electrical resistance (TEER) is a widely used as an experimental readout and a quality control assay for measuring the integrity of epithelial monolayers. However, it is difficult to compare TEER values among different Organ-on-a-Chip (OoC) devices due to considerable errors in measurements. Here we report an OoC design method focusing on the electrodes that achieves the accurate measurement of TEER values. This approach has employed an index of "sensitivity variation"calculated from current density depending on electrodes configuration and structure of cell culture chamber, allowing to identify the suitable electrodes design for TEER measurements. The validity of this approach was successfully confirmed to measure TEER by impedance spectroscopy measurements and detect changes in integrity of Caco-2 cell. The present approach will allow a comparison of TEER values across any OoC as well as real-time assessment of biological functions.

Original languageEnglish (US)
Pages (from-to)237-244
Number of pages8
JournalIEEJ Transactions on Sensors and Micromachines
Volume141
Issue number7
DOIs
StatePublished - 2021

Keywords

  • Impedance spectroscopy
  • Microfluidic device
  • Organ-on-a-Chip
  • Trans-epithelial electrical resistance (TEER)

ASJC Scopus subject areas

  • Mechanical Engineering
  • Electrical and Electronic Engineering

Fingerprint

Dive into the research topics of 'A design method of organ-on-a-chip with highly accurate measurement of trans-epithelial electrical resistance'. Together they form a unique fingerprint.

Cite this