Dynamic optical tomographic imager with optimized digital lock-in filtering

Joseph M. Lasker, James M. Masciotti, Yang Li, Chris Fong, Andreas H. Hielscher

    Research output: Chapter in Book/Report/Conference proceedingConference contribution

    Abstract

    We describe a new dynamic optical tomography system that is, unlike currently available analog instrumentation, based on digital data-acquisition and filtering techniques. At the heart of this continuous wave instrument is a digital signal processor (DSP) that collects, collates, processes, and filters the digitized data set. A digital lock-in filter that has been designed for this particular application maximizes measurement fidelity. The synchronously-timed processes are controlled by a complex programmable logic device (CPLD) that is also used in conjunction with the DSP to orchestrate data flow. Real-time data rates as high as 140Hz can be achieved. The operation of the system is implemented through a graphical user interface designed with LabVIEW software, Performance analysis shows very low system noise (~600fW RMS noise equivalent power), excellent signal precision (<0.04% - 0.2%) and long-term system stability (<1% over 40 min). A large dynamic range (~195dB) accommodates a wide scope of measurement geometries and tissue types. First experiments on tissue phantoms show that dynamic behavior is accurately captured and spatial location can be correctly tracked using this system.

    Original languageEnglish (US)
    Title of host publicationEuropean Conference on Biomedical Optics, ECBO 2007
    PublisherOptical Society of America (OSA)
    ISBN (Print)9780819467713
    DOIs
    StatePublished - 2007
    EventEuropean Conference on Biomedical Optics, ECBO 2007 - Munich, Germany
    Duration: Jun 17 2007Jun 17 2007

    Publication series

    NameOptics InfoBase Conference Papers
    ISSN (Electronic)2162-2701

    Conference

    ConferenceEuropean Conference on Biomedical Optics, ECBO 2007
    CountryGermany
    CityMunich
    Period6/17/076/17/07

    Keywords

    • Digital lock-in detection
    • Dynamic imaging
    • Instrumentation
    • Optical tomography

    ASJC Scopus subject areas

    • Instrumentation
    • Atomic and Molecular Physics, and Optics

    Fingerprint Dive into the research topics of 'Dynamic optical tomographic imager with optimized digital lock-in filtering'. Together they form a unique fingerprint.

    Cite this