T1 - Individual A-scan signal normalization between two spectral domain optical coherence tomography devices
AU - Chen, Chieh Li
AU - Ishikawa, Hiroshi
AU - Wollstein, Gadi
AU - Ling, Yun
AU - Bilonick, Richard A.
AU - Kagemann, Larry
AU - Sigal, Ian A.
AU - Schuman, Joel S.
PY - 2013/4/23
Y1 - 2013/4/23
N2 - PURPOSE. We developed a method to normalize optical coherence tomography (OCT) signal profiles from two spectral-domain (SD) OCT devices so that the comparability between devices increases. METHODS. We scanned 21 eyes from 14 healthy and 7 glaucoma subjects with two SD-OCT devices on the same day, with equivalent cube scan patterns centered on the fovea (Cirrus HD-OCT and RTVue). Foveola positions were selected manually and used as the center for registration of the corresponding images. A-scan signals were sampled 1.8 mm from the foveola in the temporal, superior, nasal, and inferior quadrants. After oversampling and rescaling RTVue data along the Z-axis to match the corresponding Cirrus data format, speckle noise reduction and amplitude normalization were applied. For comparison between normalized A-scan profiles, mean absolute difference in amplitude in percentage was measured at each sampling point. As a reference, the mean absolute difference between two Cirrus scans on the same eye also was measured. RESULTS. The mean residual of the A-scan profile amplitude was reduced significantly after signal normalization (12.7% vs. 6.2%, P < 0.0001, paired t-test). All four quadrants also showed statistically significant reduction (all P < 0.0001). Mean absolute difference after normalization was smaller than the one between two Cirrus scans. No performance difference was detected between health and glaucomatous eyes. CONCLUSIONS. The reported signal normalization method successfully reduced the A-scan profile differences between two SD-OCT devices. This signal normalization processing may improve the direct comparability of OCT image analysis and measurement on various devices.
AB - PURPOSE. We developed a method to normalize optical coherence tomography (OCT) signal profiles from two spectral-domain (SD) OCT devices so that the comparability between devices increases. METHODS. We scanned 21 eyes from 14 healthy and 7 glaucoma subjects with two SD-OCT devices on the same day, with equivalent cube scan patterns centered on the fovea (Cirrus HD-OCT and RTVue). Foveola positions were selected manually and used as the center for registration of the corresponding images. A-scan signals were sampled 1.8 mm from the foveola in the temporal, superior, nasal, and inferior quadrants. After oversampling and rescaling RTVue data along the Z-axis to match the corresponding Cirrus data format, speckle noise reduction and amplitude normalization were applied. For comparison between normalized A-scan profiles, mean absolute difference in amplitude in percentage was measured at each sampling point. As a reference, the mean absolute difference between two Cirrus scans on the same eye also was measured. RESULTS. The mean residual of the A-scan profile amplitude was reduced significantly after signal normalization (12.7% vs. 6.2%, P < 0.0001, paired t-test). All four quadrants also showed statistically significant reduction (all P < 0.0001). Mean absolute difference after normalization was smaller than the one between two Cirrus scans. No performance difference was detected between health and glaucomatous eyes. CONCLUSIONS. The reported signal normalization method successfully reduced the A-scan profile differences between two SD-OCT devices. This signal normalization processing may improve the direct comparability of OCT image analysis and measurement on various devices.
KW - Comparability
KW - Image analysis
KW - Optical coherence tomography
UR - http://www.scopus.com/inward/record.url?scp=84878162596&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84878162596&partnerID=8YFLogxK
U2 - 10.1167/iovs.12-11484
DO - 10.1167/iovs.12-11484
M3 - Article
C2 - 23611992
AN - SCOPUS:84878162596
SN - 0146-0404
VL - 54
SP - 3463
EP - 3471
JO - Investigative Ophthalmology and Visual Science
JF - Investigative Ophthalmology and Visual Science
IS - 5
ER -