TY - JOUR
T1 - Mechanical testing of implant-supported anterior crowns with different implant/abutment connections
AU - Almeida, Erika O.
AU - Freitas, Amilcar C.
AU - Bonfante, Estevam A.
AU - Marotta, Leonard
AU - Silva, Nelson R F A
AU - Coelho, Paulo G.
PY - 2013
Y1 - 2013
N2 - Purpose: This study evaluated the reliability and failure modes of anterior implants with internal-hexagon (IH), external-hexagon (EH), or Morse taper (MT) implant-abutment interface designs. The postulated hypothesis was that the different implant-abutment connections would result in different reliability and failure modes when subjected to step-stress accelerated life testing (SSALT) in water. Materials and Methods: Sixty-three dental implants (4 × 10 mm) were divided into three groups (n = 21 each) according to connection type: EH, IH, or MT. Commercially pure titanium abutments were screwed to the implants, and standardized maxillary central incisor metallic crowns were cemented and subjected to SSALT in water. The probability of failure versus number of cycles (95% two-sided confidence intervals) was calculated and plotted using a power-law relationship for damage accumulation. Reliability for a mission of 50,000 cycles at 150 N (90% two-sided confidence intervals) was calculated. Polarized-light and scanning electron microscopes were used for failure analyses. Results: The beta values (confidence intervals) derived from use-level probability Weibull calculation were 3.34 (2.22 to 5.00), 1.72 (1.14 to 2.58), and 1.05 (0.60 to 1.83) for groups EH, IH, and MT, respectively, indicating that fatigue was an accelerating factor for all groups. Reliability was significantly different between groups: 99% for MT, 96% for IH, and 31% for EH. Failure modes differed; EH presented abutment screw fracture, IH showed abutment screw and implant fractures, and MT displayed abutment and abutment screw bending or fracture. Conclusions: The postulated hypothesis that different implant-abutment connections to support anterior single-unit replacements would result in different reliability and failure modes when subjected to SSALT was accepted.
AB - Purpose: This study evaluated the reliability and failure modes of anterior implants with internal-hexagon (IH), external-hexagon (EH), or Morse taper (MT) implant-abutment interface designs. The postulated hypothesis was that the different implant-abutment connections would result in different reliability and failure modes when subjected to step-stress accelerated life testing (SSALT) in water. Materials and Methods: Sixty-three dental implants (4 × 10 mm) were divided into three groups (n = 21 each) according to connection type: EH, IH, or MT. Commercially pure titanium abutments were screwed to the implants, and standardized maxillary central incisor metallic crowns were cemented and subjected to SSALT in water. The probability of failure versus number of cycles (95% two-sided confidence intervals) was calculated and plotted using a power-law relationship for damage accumulation. Reliability for a mission of 50,000 cycles at 150 N (90% two-sided confidence intervals) was calculated. Polarized-light and scanning electron microscopes were used for failure analyses. Results: The beta values (confidence intervals) derived from use-level probability Weibull calculation were 3.34 (2.22 to 5.00), 1.72 (1.14 to 2.58), and 1.05 (0.60 to 1.83) for groups EH, IH, and MT, respectively, indicating that fatigue was an accelerating factor for all groups. Reliability was significantly different between groups: 99% for MT, 96% for IH, and 31% for EH. Failure modes differed; EH presented abutment screw fracture, IH showed abutment screw and implant fractures, and MT displayed abutment and abutment screw bending or fracture. Conclusions: The postulated hypothesis that different implant-abutment connections to support anterior single-unit replacements would result in different reliability and failure modes when subjected to SSALT was accepted.
KW - Cementation, dental implants
KW - Fractography
KW - Implant-supported prostheses
KW - Reliability
KW - Step-stress accelerated life testing
UR - http://www.scopus.com/inward/record.url?scp=84886373412&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84886373412&partnerID=8YFLogxK
U2 - 10.11607/jomi.2443
DO - 10.11607/jomi.2443
M3 - Article
C2 - 23377054
AN - SCOPUS:84886373412
SN - 0882-2786
VL - 28
SP - 103
EP - 108
JO - International Journal of Oral and Maxillofacial Implants
JF - International Journal of Oral and Maxillofacial Implants
IS - 1