T1 - Reliability and failure modes of narrow implant systems
AU - Hirata, Ronaldo
AU - Bonfante, Estevam A.
AU - Anchieta, Rodolfo B.
AU - Machado, Lucas S.
AU - Freitas, Gileade
AU - Fardin, Vinicius P.
AU - Tovar, Nick
AU - Coelho, Paulo G.
N1 - Publisher Copyright:
© 2015, Springer-Verlag Berlin Heidelberg.
PY - 2016/9/1
Y1 - 2016/9/1
N2 - Objectives: Narrow implants are indicated in areas of limited bone width or when grafting is nonviable. However, the reduction of implant diameter may compromise their performance. This study evaluated the reliability of several narrow implant systems under fatigue, after restored with single-unit crowns. Materials and methods: Narrow implant systems were divided (n = 18 each), as follows: Astra (ASC); BioHorizons (BSC); Straumann Roxolid (SNC), Intra-Lock (IMC), and Intra-Lock one-piece abutment (ILO). Maxillary central incisor crowns were cemented and subjected to step-stress accelerated life testing in water. Use level probability Weibull curves and reliability for a mission of 100,000 cycles at 130- and 180-N loads (90 % two-sided confidence intervals) were calculated. Scanning electron microscopy was used for fractography. Results: Reliability for 100,000 cycles at 130 N was ∼99 % in group ASC, ∼99 % in BSC, ∼96 % in SNC, ∼99 % in IMC, and ∼100 % in ILO. At 180 N, reliability of ∼34 % resulted for the ASC group, ∼91 % for BSC, ∼53 % for SNC, ∼70 % for IMC, and ∼99 % for ILO. Abutment screw fracture was the main failure mode for all groups. Conclusions: Reliability was not different between systems for 100,000 cycles at the 130-N load. A significant decrease was observed at the 180-N load for ASC, SNC, and IMC, whereas it was maintained for BSC and ILO. Clinical relevance: The investigated narrow implants presented mechanical performance under fatigue that suggests their safe use as single crowns in the anterior region.
AB - Objectives: Narrow implants are indicated in areas of limited bone width or when grafting is nonviable. However, the reduction of implant diameter may compromise their performance. This study evaluated the reliability of several narrow implant systems under fatigue, after restored with single-unit crowns. Materials and methods: Narrow implant systems were divided (n = 18 each), as follows: Astra (ASC); BioHorizons (BSC); Straumann Roxolid (SNC), Intra-Lock (IMC), and Intra-Lock one-piece abutment (ILO). Maxillary central incisor crowns were cemented and subjected to step-stress accelerated life testing in water. Use level probability Weibull curves and reliability for a mission of 100,000 cycles at 130- and 180-N loads (90 % two-sided confidence intervals) were calculated. Scanning electron microscopy was used for fractography. Results: Reliability for 100,000 cycles at 130 N was ∼99 % in group ASC, ∼99 % in BSC, ∼96 % in SNC, ∼99 % in IMC, and ∼100 % in ILO. At 180 N, reliability of ∼34 % resulted for the ASC group, ∼91 % for BSC, ∼53 % for SNC, ∼70 % for IMC, and ∼99 % for ILO. Abutment screw fracture was the main failure mode for all groups. Conclusions: Reliability was not different between systems for 100,000 cycles at the 130-N load. A significant decrease was observed at the 180-N load for ASC, SNC, and IMC, whereas it was maintained for BSC and ILO. Clinical relevance: The investigated narrow implants presented mechanical performance under fatigue that suggests their safe use as single crowns in the anterior region.
KW - Biomechanics
KW - Dental implants
KW - Fatigue
KW - Reliability
KW - Weibull
UR - http://www.scopus.com/inward/record.url?scp=84946112222&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84946112222&partnerID=8YFLogxK
U2 - 10.1007/s00784-015-1636-8
DO - 10.1007/s00784-015-1636-8
M3 - Article
C2 - 26526325
AN - SCOPUS:84946112222
SN - 1432-6981
VL - 20
SP - 1505
EP - 1513
JO - Clinical Oral Investigations
JF - Clinical Oral Investigations
IS - 7
ER -