Abstract
Fast, 3D radio-frequency transmit field (B1) mapping is important for parallel transmission, spatially selective pulse design and quantitative MRI applications. It has been shown that actual flip angle imaging - two interleaved spoiled gradient recalled echo images acquired in steady state with two very short time delays (TR1, TR2) - is an attractive method of B1 mapping. Herein, we describe the TROMBONE method that efficiently integrates actual flip angle imaging with EPI imaging, alleviates very short TR requirement of actual flip angle imaging and through their synergy yields up to 16 times higher precision in B1 estimation in the same experimental time. High precision of TROMBONE can be traded for faster scans. The map of B1 reconstructed from the ratio of intensities of two images is insensitive to longitudinal relaxation time (T 1) in the physiologically relevant range. A table of the optimal acquisition protocol parameters for various target experimental conditions is provided.
Original language | English (US) |
---|---|
Pages (from-to) | 483-491 |
Number of pages | 9 |
Journal | Magnetic resonance in medicine |
Volume | 66 |
Issue number | 2 |
DOIs | |
State | Published - Aug 2011 |
Keywords
- RF inhomogeneity
- actual flip angle imaging
- fast volumetric imaging
- high-field MRI
- optimization methods
- parallel transmission
- quantitative MRI
ASJC Scopus subject areas
- Radiology Nuclear Medicine and imaging